Sale!

YOKOGAWA ZR22G-100-S-C-E-T-Q-E-A Split-type Zirconia Oxygen Analyzer Detector

Brand: YOKOGAWA

Model number:ZR22G-100-S-C-E-T-Q-E-A

Colour:new                                    Warranty: 12 months

Lead Time:3-day working day   Country of origin: USA

Price: Please contact us         Product weight:0.7kg

Category: SKU: ZR22G-100-S-C-E-T-Q-E-A Brand:

Description

YOKOGAWA ZR22G-100-S-C-E-T-Q-E-A Split-type Zirconia Oxygen Analyzer Detector

Product Dimensions:

The Yokogawa ZR22G-100-S-C-E-T-Q-E-A is a separate-type Zirconia Oxygen Analyzer probe. It is part of the ADMAG ZR series, renowned for its durability and precision in monitoring oxygen concentration in combustion flue gases.

Below is a detailed breakdown of this specific model.

1. Product Description

The ZR22G is a general-purpose probe that uses a Zirconia sensor to measure oxygen concentration. It is designed to be inserted directly into a flue or furnace. This “separate-type” system means the probe is installed at the measurement point and connected via a cable to a remote converter (usually the ZR402G).

2. Product Parameters (Based on Model Code)

The suffix code -100-S-C-E-T-Q-E-A defines the specific configuration:

Code Specification Detail
ZR22G Separate type Zirconia Oxygen Analyzer Probe
-100 Length: 1.0 meter
-S Tube Material: SUS 316 Stainless Steel
-C Connection: ANSI Class 150 4-RF (or equivalent flange)
-E Reference Air Inlet: 1/2 NPT Female
-T Gas Outlet: 1/4 NPT Female
-Q Sensor Type: High-reliability Zirconia sensor with heater
-E Instruction Manual: English
-A Special Requirement: Standard version / No specific customization

Technical Specifications:

  • Measurement Range: 0.01% to 100% $O_2$

  • Process Temperature: Up to 700°C (standard)

  • Sample Gas Pressure: -5 to +5 kPa

  • Repeatability: $\pm 0.5\%$ of maximum scale

  • Response Time: 90% response within 5 seconds

3. Product Advantages

  • In-situ Measurement: Eliminates the need for complex sampling systems, reducing maintenance and lag time.

  • Field-Replaceable Sensor: The Zirconia cell can be replaced easily on-site without disposing of the entire probe.

  • High Accuracy: Features a high-stability heater control to ensure precise readings even during process temperature fluctuations.

  • Long Life: The heater and sensor assembly are built to withstand harsh, corrosive combustion environments.

4. Product Applications

This probe is primarily used for combustion control and environmental monitoring in:

  • Power Plants: Optimizing fuel-to-air ratios in large boilers.

  • Refineries: Monitoring process heaters and crackers.

  • Iron & Steel: Reheating furnaces and soaking pits.

  • Incinerators: Ensuring complete combustion and emission compliance.

5. Important Notes & Precautions

  • Thermal Shock: Avoid rapid cooling of the sensor. Do not use water or wet air to clean a hot probe.

  • Mounting Orientation: Ideally installed horizontally or vertically downward. Avoid upward installation where condensate can collect on the sensor.

  • Calibration: Periodic calibration with “Standard Gas” is required to maintain accuracy. Ensure the calibration gas lines are leak-free.

  • Safety: The probe operates at high temperatures (the internal heater reaches roughly 750°C). Use caution during handling and maintenance.

Shipping Port: Xiamen, China

Payment: Bank of Chicago, Bank of Singapore

Express cooperation: fedex, DHL, UPS and your express account

Service: Professional Sales provides 24 hours /7 days online service

Please make an inquiry for more models:

YOKOGAWA CP451
YOKOGAWA LR4220E
YOKOGAWA ST6A
YOKOGAWA MS31*A
YOKOGAWA MX6*A
YOKOGAWA NC4*B
YOKOGAWA PS31*A
YOKOGAWA MAC2*B
YOKOGAWA FCA4*B
YOKOGAWA VM1*D
YOKOGAWA LC82*A
YOKOGAWA 2150S
YOKOGAWA ST6*A
YOKOGAWA C2L315601-1111
YOKOGAWA DM1A-050G
YOKOGAWA MW100-E-1R
YOKOGAWA AV8C-A-4-EA/4P
YOKOGAWA DT300-31
YOKOGAWA EB511-10
YOKOGAWA PS65*A
YOKOGAWA AAI543-S03
YOKOGAWA AFV10D-S41201
YOKOGAWA CA71
YOKOGAWA ER5*B
YOKOGAWA ET5*B
YOKOGAWA ET5*C
YOKOGAWA SDV144-S53
YOKOGAWA SDV541-S53
YOKOGAWA STB4D-10
YOKOGAWA VI701-S1